欧美肥妇BWBWBWBXX小说,巜年轻的公和我做愛,男生裸睡勃起流出液体,少妇又紧又色又爽又刺激视频

本站熱搜:氧化鋯氧量分析儀,盤裝式氧量分析儀,電磁流量計,WSS雙金屬溫度計
  • 技術(shù)文章ARTICLE

    您當(dāng)前的位置:首頁 > 技術(shù)文章 > Common faults and causes of zirconia oxygen analyzers

    Common faults and causes of zirconia oxygen analyzers

    發(fā)布時間: 2024-12-13  點擊次數(shù): 179次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


產(chǎn)品中心 Products
在線客服 聯(lián)系方式 二維碼

服務(wù)熱線

0550-2409223

掃一掃,關(guān)注我們

皖公網(wǎng)安備 34118102000850號

小舞屈辱打开双腿自慰出白浆| 粗大抽搐白浊H高干H| 一性一交一伦一片A片M3U8| 丰满多毛的大隂户毛茸茸| 紫黑硕大撕裂高h| 欧美激情视频| 丰满熟妇岳av无码区hd | 国产人与禽zoz0性伦| 欧美性受xxxx| 日韩精品人妻一区二区中文八零 | 交换俱乐部娇妻奶呻吟啊视频| 免费看裸体美女脱了衣服露视频胸 | 成人视频网站| 强开美妇后菊哀嚎哭叫视频| 真人做爰高潮全过程毛片| 亚洲精品久久久久一区二区| 日韩欧美亚洲国产精品字幕久久久| 朝鲜女人毛多水多又紧又滑| 高h粗口调教羞辱sm文女王| 思思久久精品在热线热| 婷婷综合另类小说色区| 日本欧美久久久久免费播放网 | 国产av在线观看| 国产真人毛片爱做A级毛| 富婆被大肉楱征服小说| 女人张开腿让男人添| 古代翁妇乱高h辣文| 快被强欲学长c晕了男男小说| 堕落的人妻1―10雨柔| 午夜成人影院| 亚洲精品国产精品乱码不99| 无码人妻精品一区二区蜜桃网站| 清区二三区国产好的精华液| NBA直播在线观看免费| 绿帽娇妻肚子被灌满精怀孕| 办公室撕开奶罩揉吮奶漫画| 日本16岁RAPPER| 成人性生交大片免费看中文| 精品人妻无码一区二区三区淑枝| 少妇一夜三次一区二区| 国产xxx69麻豆国语对白|